Metrohm IC Net 2.2 User Manual
Page 31

3.4 Global settings
IC Net 2.2
23
togram is saved.
If this option is disabled, the directory is
opened where the last opened chroma-
togram is saved.
21 CFR Part 11 On
Switch on/off global 21 CFR Part 11 set-
tings. Is automatically activated and cannot
be altered, if IC Net Compliant is installed.
The following parameter settings are auto-
matically set:
Data file overwrite
= Never
If method changed
= Ask to save
If disk method is newer = Ask to overwrite
User comments
= Require Comment
on changes enabled
GLP On
= enabled
GLP On
Switch on/off global GLP settings. The fol-
lowing parameter settings are automatically
set:
Data file overwrite
= Never
If method changed
= Ask to save
If disk method is newer = Ask to overwrite
System face
Stacked recorder icons
If this option is enabled, only one recorder
icon with different tabs appears in the sys-
tem window if several data recorders are
installed.
OK, Apply buttons mean
Save object to file
If this option is enabled, the
button
in system settings windows is replaced by
the
button. The system settings are
saved if the
or
button is
clicked.
Chromatogram units
Units for chromatograms:
Flow
Unit for flow rate:
µ
L/min
,
mL/min
Pressure
Unit for pressure:
MPa
,
psi
,
bar
,
atm
Print via print spooler
Switch on/off printing via print spooler.
Switch off this option if problems with print-
ing occur.
Default colors for chromatographic win-
dows (details see section 8.3.1).