Metrohm IC Net 2.2 User Manual
Page 68
![background image](/manuals/586876/68/background.png)
6 Devices
IC Net 2.2
60
Pmax
0.0 ... 25.0 MPa
Set maximum pressure limit for
the high-pressure pump con-
nected to the 732 IC Detector to
the desired value.
Pmin
0.0 ... 25.0 MPa
Set minimum pressure limit for the
high-pressure pump connected to
the 732 IC Detector to the desired
value.
Remote configuration
On the
Remote configuration
subtab user-defined remote com-
mands can be defined, which can be inserted into a time pro-
gram after being activated with
.
Name (1
st
column)
User-definable name of the remote com-
mand (e.g.
Start_766
).
Remote command (2
nd
column)
Setting the remote output lines 1
…
8.
0
(line off, inactive, open)
1
(line on, active, 0 V)
*
(leave line in current status)
For entry of the first value, enter
1
,
0
or
*
.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter
1
,
0
or
*
.
Add new remote command.
Delete selected remote command.
Activate the defined remote commands for
insertion into the time program.
Links
The
Links
tab of the
732 IC Detector
is used for COM port selection
and settings (details see section 5.2.4 Links).