4.7 Results

763 PVC Thermomat

87

Parameters

Parameters

tab of the

Determination and method data

window with the dis-

play of the measuring parameters used in the determination.

Temperature

Set temperature of sample.

Delta T

Temperature correction: Correction value for

increasing the block temperature in order to

reach the set temperature in the sample.

Gas flow

Gas flow through sample.

Start options

Start delay

Delay time until start of measurement.

Start mode

Mode for starting the channels:

per channel

All channels have been started individually.

per 2 channels

Each 2 channels have been started together.

per block

4 channels in one block have been started to-

gether.

Stop criteria

Time

Time until automatic termination of determina-

tion.

Conductivity

Conductivity limit. When this value is reached

the determination is automatically terminated.

Endpoint(s)

Automatic termination of the determination

when each active channel has reached all end-

points.