beautypg.com

Metrohm 763 PVC Thermomat Manual User Manual

Page 107

background image

4 Operation

763 PVC Thermomat

100

With the

Copy

option in the context-sensitive menu, which is opened by

clicking on the graph with the right-hand mouse key, the displayed ex-

trapolation graph can be copied into the clipboard.

Extrapolation results

The results of the extrapolation calculation are displayed beneath the

graph:

Formula

Exponential equation for calculating the ex-

trapolated time for the target temperature from

the measuring values.

A

Calculated coefficient

A

of the exponential

equation.

B

Calculated coefficient

B

of the exponential

equation.

r^2

Calculated coefficient of correlation. If this value

is below the

limit

(see below) set by the admin-

istrator then a warning appears when the ex-

trapolation window is opened.

Time

Result of the extrapolation for the target tem-

perature in h and years.

Target temperature

Target temperature for the extrapolation.

Number of determinations

Number of determinations used for the ex-

trapolation.

Standard factor

Value determined from the regression line

which gives the factor by which the time

changes per 10 °C. This value can be entered

in the equation for determination of the

Standard

time

as a

Factor

(see section 4.5.2).

Opens the

Program settings

window with the

Ex-

trapolation

tab for the entry of parameters for

extrapolation (see below).

Updates the display of graph and results after

an alteration in the selection of the determina-

tions in the determination overview.

Shows only the measured values in the graph.

Print extrapolation results

763 PVC THERMOMAT RESULTS / File / Print

If the

Extrapolation

window is open then this menu item prints out the

extrapolation graphs (normal and detail), the extrapolation results and

an overview of the determinations used for the extrapolation with the

ID

1

,

ID 2

,

Temperature

and

Results

fields.