4 Operation
763 PVC Thermomat
100
With the
Copy
option in the context-sensitive menu, which is opened by
clicking on the graph with the right-hand mouse key, the displayed ex-
trapolation graph can be copied into the clipboard.
Extrapolation results
The results of the extrapolation calculation are displayed beneath the
graph:
Formula
Exponential equation for calculating the ex-
trapolated time for the target temperature from
the measuring values.
A
Calculated coefficient
A
of the exponential
equation.
B
Calculated coefficient
B
of the exponential
equation.
r^2
Calculated coefficient of correlation. If this value
is below the
limit
(see below) set by the admin-
istrator then a warning appears when the ex-
trapolation window is opened.
Time
Result of the extrapolation for the target tem-
perature in h and years.
Target temperature
Target temperature for the extrapolation.
Number of determinations
Number of determinations used for the ex-
trapolation.
Standard factor
Value determined from the regression line
which gives the factor by which the time
changes per 10 °C. This value can be entered
in the equation for determination of the
Standard
time
as a
Factor
(see section 4.5.2).
Opens the
Program settings
window with the
Ex-
trapolation
tab for the entry of parameters for
extrapolation (see below).
Updates the display of graph and results after
an alteration in the selection of the determina-
tions in the determination overview.
Shows only the measured values in the graph.
Print extrapolation results
763 PVC THERMOMAT RESULTS / File / Print
If the
Extrapolation
window is open then this menu item prints out the
extrapolation graphs (normal and detail), the extrapolation results and
an overview of the determinations used for the extrapolation with the
ID
1
,
ID 2
,
Temperature
and
Results
fields.