beautypg.com

Metrohm 763 PVC Thermomat Manual User Manual

Page 69

background image

4 Operation

763 PVC Thermomat

62

Standards

Tab

Standards

of the

Method parameters

window with parameters for cal-

culating the standard time.

Calculate standard time

Switches the calculation of the standard time

according to the following equation on/off:

Standard time = Factor

(Temperature - Target temperature)/10

×× Induction time

Formula parameters

Factor

Parameter for calculating the standard time

which gives the factor by which the time

changes by 10 °C. This value can be deter-

mined by the extrapolation (see section 4.7.8)

of results obtained at different temperatures.

Target temperature

Target temperature for calculating the standard

time.