4.7 Results

763 PVC Thermomat

101

Settings for extrapolation

763 PVC THERMOMAT RESULTS / Options / Settings / Extrapolation

Extrapolation

tab of the

Program settings

window (see section 4.7.7) with

settings for extrapolation.

Induction time

Extrapolation with the results for the

induction

time

.

Stability time

Extrapolation with the results for the

stability

time

.

Formula 1..3 ...

Extrapolation with the

Formula results 1...3

.

Target temperature

Temperature to which the selected result is to

be converted with the help of the regression

line obtained from the extrapolation.

Parameters for extrapolation

763 PVC THERMOMAT RESULTS / Options / Administrator / Extrapolation parameters

The

Extrapolation parameters

window opens.

Correlation limit

Limit for the correlation coefficient

r

2

determined

in the extrapolation. If this value is below the

limit a warning will be given.

This menu item is reserved for the

Administrator

.