4.6 Determinations

763 PVC Thermomat

69

4.6.4

Carry out determinations

Select method

The determination method is selected in the

Method

field in the control

window.

Different methods with different temperatures can be selected for Block

A and Block B. In contrast, the gas flow which is defined in the method

loaded for Block A also applies to Block B.

By clicking on the method parameters (see section 4.5.2) can be

altered as required.

Switch on heating

The heating of the two blocks A and B to the temperature set in the

method can be switched on individually for each block. This is done by

clicking on the following button.

As soon as the heating is switched on the color of the frame of this

button switches to red. At the same time the temperature display 99

"TEMPERATURE" on the 763 PVC Thermomat starts to blink for the

selected block.

If this button is clicked on again during the heating-up phase then the

heating will be switched off.

The actual temperature is shown digitally beside the button. A bar be-

low shows the absolute variation of the actual temperature from the tar-

get temperature; the bar length corresponds to 50 °C.

When the selected temperature has been reached the temperature dis-

play 99 "TEMPERATURE" stops blinking and the LED remains lit up. At

the same time the color of the frame of the button switches to green.

The block is now ready for receiving the samples and starting the de-

termination.