2 Installation

763 PVC Thermomat

14

3 Change fuses

Remove old fuses carefully from fuse holder 14

14 and replace

them with two new fuses which are suitable for the set mains

voltage, type TH (slow-blow, with high switching capacity):

115 V

4 A (TH)

Metrohm-No. U.600.0022

230 V

2 A (TH)

Metrohm-No. U.600.0107

4 Install fuse holder

Push fuse holder 14

14 back into the instrument until it clicks into

position.

2.3.3

Mains cable and mains connection

Mains cable

The instrument is supplied with one of three mains cables

6.2122.020 with plug SEV 12 (Switzerland,

)

6.2122.040 with plug CEE(7), VII (Germany,

)

6.2133.070 with plug NEMA 5-15 (USA,

)

which are three-cored and fitted with a plug with an earthing pin. If a

different plug has to be fitted, the yellow/green lead (IEC standard)

must be connected to protective earth (protection class 1).

Any break in the earthing inside or outside the instrument can make it

a hazard!

Mains connection

Plug the mains cable into mains connection plug 13

13 of the 763 PVC

Thermomat (see Fig. 3).

2.3.4

Switching the instrument on/off

The 763 PVC Thermomat is switched on and off using mains switch 12

12.

When the instrument is switched on, the mains pilot lamp 77 "ON" (see

Fig. 2) lights up.