4.8 GLP

763 PVC Thermomat

123

GLP conductivity test

763 PVC THERMOMAT CONTROL / Tools / GLP test / Conductivity

This menu item is used to start the automatic GLP test for conductivity

measurement. This test can either be carried out with a standard solu-

tion with known conductivity or with the 6.2109.030 Test resistor which

is included in the 6.5616.000 GLP test set available from Metrohm as

an option. The test resistor must be connected to the electrode con-

nections 20

20 of the 763 PVC Thermomat.

In the GLP test a separate window opens which guides you through the

test step-by-step. The next step is reached with

,

the previous

step with

and the test is aborted with

.

The cell constants of the conductivity cells have to be determined

before carrying out the GLP conductivity test with standard solutions

for the first time (see section 4.4.1).

Proceed as follows to carry out the GLP conductivity test:

1 Select instrument

Select unit 1...4 in the control window for which the GLP

conductivity test has to be carried out.

If the GLP test is carried out with the test resistor carry on

from step 2, otherwise continue with step 3.

2 Carry out GLP test with test resistor

Click on

763 PVC THERMOMAT CONTROL / Tools / GLP test /

Conductivity

and click on

in the GLP test window.