4 Operation

763 PVC Thermomat

56

The

Method parameters

window contains the following buttons:

Saves method under the same name.

Saves method under a new name; the old

method is retained.

Prints out method report.

Closes method parameter window without

saving the modified parameters.

Parameters

Tab

Parameters

of the

Method parameters

window with measuring pa-

rameters.

Temperature

Set temperature of the sample.

Range:

50...220 °C

Delta T

Temperature correction: Correction value for

increasing the block temperature in order to

reach the set temperature in the sample. This

value is determined by measuring the variation

of the actual sample temperature from the

block temperature using a calibrated external

temperature sensor (see section 4.4.2).

Range:

-9.9...9.9 °C