Metrohm 763 PVC Thermomat Manual User Manual

Page 57

background image

4 Operation

763 PVC Thermomat

50

4.4.2

Determination of Delta T

The temperature correction Delta T describes the variation of the

actual sample temperature from the temperature in the heating block

and is a parameter of the method (see section 4.5.2). It can be auto-

matically determined with the 6.1111.010 Calibrated external tempera-

ture sensor (part of the 6.5616.000 GLP test set available from Metrohm

as an option). Since the direct temperature measurement in PVC sam-

ples is impossible, silicone oil must be used as sample instead of PVC.

Preparations

Fig. 5 shows in detail how to install the accessories for Delta T determi-

nation. Proceed as follows:

1 Prepare reaction vessel cover

Install gas inlet tube 35

35 (part of the 6.5616.000 GLP test set,

length = 148 mm) on the reaction vessel cover 22 (see section

2.2.2).

Fasten distance piece 37

37 to the gas inlet tube 35

35 so that it is

located at a distance of approx. 2 cm from the upper end of

the reaction vessel.

Fasten distance piece 37

37 to the gas inlet tube 35

35 so that it is

located at a distance of approx. 5 cm from the lower end of

the reaction vessel.

Insert temperature sensor 36

36 from above in opening 31

31 of the

reaction vessel cover 22 and fasten it in the corresponding

openings of the distance pieces 37

37.

2 Prepare reaction vessel

Fill reaction vessel 11 with 5 g silicone oil (e.g. FLUKA

85409).

Place the reaction vessel cover 22 with temperature sensor 36

36

on the reaction vessel 11.

Slide the temperature sensor 36

36 right down (the sensor must

touch the base of the vessel).

3 Insert and connect reaction vessel

Place the prepared reaction vessel 11 in channel 2 or 3 of the

heating block.

Attach one end of the brown connecting tubing 66 at connec-

tion nipple 30

30 on the reaction vessel cover 22.

Attach the other end of the brown connecting tubing 66 at the

connection 19

19 on the 763 PVC Thermomat (see Fig. 3).

4 Modify method parameters

In the control window select the required

Method

for the

heating block for which the temperature correction

Delta T

is

to be determined.