beautypg.com

Metrohm 763 PVC Thermomat Manual User Manual

Page 74

background image

4.6 Determinations

763 PVC Thermomat

67

When the instrument is not in use always close the openings

with stoppers.

2 Measuring vessels and accessories

After the contents have been discarded clean used meas-

uring vessels 44 several times with alcohol and distilled water

(do not use acetone!).

Rinse the measuring vessel cover 55 complete with elec-

trodes 28

28 and PTFE tube 27

27 several times with distilled water

and acetone. If necessary, remove the protective ring 29

29 for

better access to the electrodes 28

28 (see Fig. 4).

Rinse the connecting tubing 33 between reaction vessel and

measuring vessel several times with distilled water and ace-

tone.

3 Reaction vessels and accessories

It is recommended to use new reaction vessels 11 and gas

inlet tubes 32

32 for each determination. Blow out new reaction

vessels with nitrogen.

Used and not too strongly contaminated reaction vessels and

gas inlet tubes can be cleaned by immersion in boiling RBS

solution for 1 h or a similar laboratory flushing agent. They

must then be thoroughly rinsed with distilled water and ace-

tone.

RBS attacks glass. Reaction vessels and gas inlet tubes

therefore cannot be cleaned ad infinitum.

Remove the gas inlet tube 32

32 from the reaction vessel

cover 22 and rinse it several times with distilled water and

acetone. Dry the cover at 80 °C in a drying oven.

Reaction vessel covers which have not been dried properly

can falsify the results of the following determinations.

Prepare measuring vessels

The cleaned measuring vessels must always be prepared as follows

before the start of a determination (see Fig. 4):

1 Fill measuring vessels

Fill each cleaned measuring vessel 44 with 60 mL distilled

water. For analysis times >24 h more than 60 mL distilled

water must be added to take the evaporation losses of ap-

prox. 7 mL/d into account and to ensure that the electrodes

remain adequately immersed.