beautypg.com

Metrohm 763 PVC Thermomat Manual User Manual

Page 133

background image

4 Operation

763 PVC Thermomat

126

GLP gas flow test

763 PVC THERMOMAT CONTROL / Tools / GLP test / Gas flow

This menu item is used to start the automatic GLP test for gas flow

measurement. A gas flow meter is required for this test (not available

from Metrohm) so that the gas flow can be measured at each channel.
In the GLP test a separate window opens which guides you through the

test step-by-step. The next step is reached with

,

the previous

step with

and the test is aborted with

.

Proceed as follows to carry out the GLP gas flow test:

1 Select instrument

Select unit 1...4 in the control window for which the GLP gas

flow test has to be carried out.

2 Start GLP test

Click on

763 PVC THERMOMAT CONTROL / Tools / GLP test / Gas

flow

and click on

in the GLP test window.

Gas flow rate

Gas flow in L/h.

Enter the

Gas flow rate

and click on

.