4 Operation

763 PVC Thermomat

122

Ext. temperature

The temperature measured by the external

temperature sensor.

Measuring time

The elapsed time since the start of the tem-

perature measurement.

When the measurement has been carried out you are re-

quested to insert the external temperature sensor together

with the GLP insert in the next selected channel and to start

the measurement. The change must be carried out quickly.

Carry out the GLP temperature measurement for each se-

lected channel in sequence.

5 Confirm/cancel test results

When the last temperature measurement has been com-

pleted the

TEST RESULTS

for the measured channels are

shown.

Test passed: The variation of the measured

temperature is within the preset tolerance.

Test failed: The variation of the measured
temperature is larger than the preset tolerance.

Check the results and click on

to save the GLP test

results or on

to abort the GLP test without saving the

results.