4.8 GLP

763 PVC Thermomat

125

Temperature

Conductivity

18 °C

127

µ

S/cm

19 °C

130

µ

S/cm

20 °C

133

µ

S/cm

21 °C

136

µ

S/cm

22 °C

138

µ

S/cm

23 °C

141

µ

S/cm

24 °C

144

µ

S/cm

25 °C

147

µ

S/cm

Click on

in the GLP test window. The conductivity of

each channel is measured in sequence.

4 Confirm/cancel test results

When the last measurement has been completed the

TEST

RESULTS

(set values and measured values) for the measured

channels will be shown.

Test passed: The variation of the measured

conductivity is within the preset tolerance.

Test failed: The variation of the measured
conductivity is larger than the preset tolerance.

Check the results and click on

to save the GLP test

results or on

to abort the GLP test without saving the

results.