4 Operation

763 PVC Thermomat

120

Select the channels, for which the temperature test has to be

carried out. Enter the desired

Test duration per channel

and

click on

.

3 Insert GLP test set

Insert the temperature sensor 36

36 into the opening of the PTFE

cylinder 38

38 from above (see Fig. 6).

Position aluminium cylinder 41

41 vertically and introduce tem-

perature sensor 36

36 into the aluminium cylinder from above

until it reaches the stop.

Fasten temperature sensor 36

36 to aluminium cylinder 41

41, by

gently tightening hexagonal screw 39

39 with the help of Allen

key 40

40.

Place temperature sensor 36

36 together with PTFE and alu-

minium cylinder in the first channel selected for Block A.

Connect temperature sensor 36

36 to connection 17

17 "Pt100" on

the rear panel of the 763 PVC Thermomat and click on

to start the measurement.