4.7 Results

763 PVC Thermomat

97

The following reevaluation possibilities are offered:

Set induction time manually
If this option is switched on then the

induction time

can be set

manually. This is done by pressing the [ Ctrl ] key and at the
same time moving the red vertical line in the

Live graph

window

with pressed-down left-hand mouse key to the required lo-
cation. The new manually set induction time is displayed in the

Manual

field and in the

Live graph

window.

Set stability time manually
If this option is switched on then the alteration in conductivity

Delta Kappa

for the calculation of the

stability time

can be al-

tered manually. After alterations in the

Delta Kappa

field the

newly calculated

Stability time

is shown both in the field with

the same name and in the

Live graph

window.

Set tangents manually
If this option is switched on then tangents can be set manually
to determine the

induction time

. The induction time is then de-

fined as being the intersection point of the tangents. The tan-
gents are set as follows:

Set first tangent
Press the [ Ctrl ] key and use the left-hand mouse key to