3 Operating tutorial

763 PVC Thermomat

22

3.1.3

Carrying out determinations

When the preparations are complete the determinations can be started.

Proceed as follows:

1 Wait until the temperature has been reached

Wait until the heating block reaches the temperature defined

in the method. This condition is indicated by temperature dis-

play 99 "TEMPERATURE" which stops blinking and remains

lit up. At the same time or after a slight delay the frame color

of the

button turns green.

2 Insert reaction vessels

Place the prepared reaction vessels 11 in the heating block

openings on the 763 PVC Thermomat. Close any openings

which are not used with stoppers or empty reactions vessels

to protect against contamination.

Connect the white connecting tubing 33 fastened to the reac-

tion vessel cover 22 to the corresponding connection 22

22 on

the measuring vessel cover 55.

Screw the brown gas supply tubing 66 onto the connection

nipple 30

30 of the reaction vessel cover 22 and the correspond-

ing connection 19

19 on the 763 PVC Thermomat (see Fig. 3).

3 Start determination

In the operating field of the control window for every determi-

nation you want to start you must click on the button

.

After the start of a determination the live curve of the started

determination is shown in the operating window beside the

channel number. The actual measured value appears above

the curve.

As soon as a point of inflection is found during the measure-

ment the determination is terminated. The window with the

live curve becomes gray again and the message

finished

ap-

pears.