Metrohm 763 PVC Thermomat Manual User Manual
Page 143

6 Appendix
763 PVC Thermomat
136
6.1.3
Conductivity measurement
Sensor
6.0913.130 Conductivity cell built into
measuring vessel cover
Construction
Conductivity measurement cell with 2 steel
electrodes
Measurement principle
Alternating current measurement with 1 kHz
frequency and approx. 1.7 V amplitude
(peak to peak).
Cell constant
1.00
…
1.20
(the exact value can be manually entered or
automatically determined)
Measurement range
0
…
400
µ
S/cm
Resolution
0.1
µ
S/cm
Maximum error
±
1 %
±
0.5
µ
S/cm
6.1.4
Gas flow regulation
Gas flow supply
External supply from N
2
cylinder
Adjustable range
7
…
25 L/h
Max. variation from
set range
±
10 %
6.1.5
GLP test set
External temperature sensor
Type
Pt100 jacketed element 1.9 mm dia.,
with 4-lead technology; DIN IEC 751 class B
Calibration accuracy
0
…
90 °C:
±
0.02 °C
90
…
220 °C:
±
0.03 °C
Max. variation of measured
temperature from true value
(50
…
220 °C)
±
0.3 °C
Test resistor
Resistance
10 k
Ω
6.1.6
RS232 interface
Connector
Dsub 9-pin (male)
Function
TxD and RxD signal for connection with
software handshake
Default settings
9600 baud, 8 bit, 1 stop bit, no parity,
XON/XOFF
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)