5.2 Faults and malfunctions
763 PVC Thermomat
133
Operating status
Display
Status/Error
Rectification
ON
Instrument switched on.
ON
Instrument switched off.
Self-test following switch-on
After the instrument has been switched on an internal hardware self-test
is carried out. Each of the 7 test steps only lasts about a second and is
shown on instrument display 11
11 by a particular combination pattern of
the 4 display lamps. If there is a fault then the corresponding LEDs will
flash.
Display
Status/Error
Rectification
DEVICE
LED test. If all LEDs do not light
up during this test then an LED
is faulty.
Inform Metrohm service.
DEVICE
RAM test error.
Switch instrument off and on
again. If the error reoccurs
please contact Metrohm
service.
DEVICE
PROM check sum test error.
Switch instrument off and on
again. If the error reoccurs
please contact Metrohm
service.
DEVICE
BUSY test error.
Switch instrument off and on
again. If the error reoccurs
please contact Metrohm
service.
DEVICE
ADC test error.
Switch instrument off and on
again. If the error reoccurs
please contact Metrohm
service.
DEVICE
EEPROM test error.
Switch instrument off and on
again. If the error reoccurs
please contact Metrohm
service.
DEVICE
Incorrect mains voltage
(Variation > 10%).
Check power supply.
Gas flow display during operation
Display
Status/Error
Rectification
FLOW
Gas flow switched off.
FLOW
Gas flow control active, set gas
flow not yet reached.
FLOW
Gas flow control active, set gas
flow reached.