4.5 Methods

763 PVC Thermomat

57

If no temperature sensor is available for the

Delta T

determination, a value of +0.6 °C can be

entered for temperatures of 160 °C to 200 °C

(this value was determined using 5 g silicone oil

and a gas flow of 7 L/h nitrogen).

Gas flow

Gas flow through sample. It cannot be set in-

dividually for each block; the flow rate for both

blocks depends always on the value entered in

the method loaded for Block A.

Range:

4...15 L/h

Start options

Start delay

Delay period until start of measurement.

Range:

0...1000 min

Start mode

Mode for start and stop of the channels:

per channel

All 8 channels can be started and stopped indi-

vidually.

per 2 channels

Each 2 channels can be started and stopped

together.

per block

The 4 channels in one block can be started and

stopped together.

Stop criteria

Time

Time until automatic termination of the determi-

nation.

Range:

0.1...100 h

Conductivity

Limit value for conductivity. If this value is

reached then the determination is terminated

automatically.

Range:

1...400

µµS/cm

Endpoint(s)

Automatic termination of the determination

when each active channel has reached all end-

points.

Shut off at end of determinations

Heater

Automatically switches off the heating at the

end of the determination.

Gas flow

Automatically switches off the gas flow at the

end of the determination.

If the gas flow is automatically switched off,

the reaction vessels must be removed

immediately from the PVC Thermomat after

the end of the determination. If the reaction

vessels are not removed, the instrument can

be damaged.