Metrohm 763 PVC Thermomat Manual User Manual
Page 49

4 Operation
763 PVC Thermomat
42
763 PVC THERMOMAT CONTROL / Tools / Gas flow control / Display gas flow...
The
Gas flow display
window opens in which the actual gas flow is
shown. This function is only available if no determination is running.
4.2.5
Recording temperature
763 PVC THERMOMAT CONTROL / Tools / Temp. logging / Block A on
Switches on temperature recording for block A and external tempera-
ture sensor.
763 PVC THERMOMAT CONTROL / Tools / Temp. logging / Block A off
Switches off temperature recording for block A and external tempera-
ture sensor.
763 PVC THERMOMAT CONTROL / Tools / Temp. logging / Block B on
Switches on temperature recording for block B and external tempera-
ture sensor.
763 PVC THERMOMAT CONTROL / Tools / Temp. logging / Block B off
Switches off temperature recording for block B and external tempera-
ture sensor.
The temperature can be individually recorded at any time for each
heating block. The measured values are written in a text file which con-
tains the following values:
t(s)
Time in s
Block [°C]
Measured temperature in heating block
Ext. Sensor [°C]
Measured temperature with the external tem-
perature sensor
The text file is stored under the name
U#X-YYMMDD-HHMMSS.txt
in the
folder
Log
;
#
stands for the instrument number (1...4),
X
for the block (A
or B) and
YYMMDD-HHMMSS
for the date and time of the start of data re-
cording.