11 troubleshooting, Troubleshooting – Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual

Page 85

background image

▪▪▪▪▪▪▪

83

11

Troubleshooting

The XDS OptiProbe Analyzer is a dependable, trouble-free instrument, designed for many years of
service in your laboratory. In spite of the rugged design, problems may arise that require attention.
This guide is intended as a means of diagnosing minor problems.

There are no user-serviceable parts inside the instrument enclosure. Because of this design, we
emphasize that under no circumstances should the user attempt to open the instrument cabinet and
service any part. The components may be damaged or misaligned by handling. Hazardous voltages
may be present even with power removed from the system. Any diagnosis of internal function should
be performed using software diagnostics, not by internal inspection.

For minor problems, this guide should be consulted. While some recommendations are quite basic,
some of the suggestions may be helpful in avoiding oversights or problems. Follow the
recommendations and eliminate any possible causes listed. Recommendations are listed in logical
order of occurrence wherever possible. When all recommendations have been checked, and if the
instrument is not operating, please contact Metrohm Service Department (or your local authorized
distributor) if the problem is not solved.

Observed Problem

Recommendations

1

Instrument does not “power
up”.

1.

Verify that AC power cord is plugged in to AC

power source.

2.

Verify that the AC power cord is plugged into the

instrument AC power block.

3.

Verify that the AC power switch is turned on.

4.

Verify that AC power is available at the AC power

source, using an AC voltmeter.

5.

Check for blown fuses. If fuses are blown,

investigate and repair the cause, then replace
fuses.