This dialog box is displayed.

Next a sample scan is taken (on the
reference) and this box is displayed. This
takes about 20-30 seconds.

Wavelength Linearization is performed in two
sections, each of which takes about 45
seconds.

The user must confirm the linearization to
proceed. A dialog box will appear, and the
user should click “Yes” to proceed to the next
step.

Following Wavelength Linearization, the user
is asked to select a standard file. This is
provided on the mini-CD in the standard box.
(It may be manually copied on the computer
in the Vision directory, if desired.)

Select this file and click on “Open”.

Vision acquires a new reference scan.

Vision requests that the XC-1300 WST3Wxxx
standard be placed into the calibration
fixture.