After each standard is run, Vision plots the
comparison for each wavelength area as
shown. Tolerances are automatically applied,
and a “Pass” or “Fail” indication is given.

Continue to follow the on-screen prompts for
each standard. Vision requests the T02, T03,
T04, and T05 standards.

Progress is shown on the screen. The split
screen has been “opened” to the left for a
better view of the data. The tabulated data is
shown below, as a typical example.

When Vision has completed the test, the tabulated results
may be printed. Test results are also stored in the
Diagnostic Database for later recall.

When the test is complete, click “Close Report”.