Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual

Page 51

background image

▪▪▪▪▪▪▪

49

Vision may open to the Diagnostics “split
screen” to show the acquisition process
for each standard, when loaded.

No spectrum is shown while the
reference scan is performed. The red
progress bar at the bottom of the screen
shows progress with the reference scan.

Vision requests the T013xxxx standard
from the set.

Select the requested standard from the
set. Labels on the top identify each
standard.

Note the mini-CD that contains the
“Standards File”. This file is used during
IPV Setup.

A different file is generated during IPV
Setup, which will be used in Photometric
Test.

Insert the requested standard into the
calibration fixture as shown, close the lid,
and click “OK” to proceed.