Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual

Page 10

background image

8

▪▪▪▪▪▪▪

standard RJ-45 type cable, such as CDW #074092, available from CDW Computer Centers, Inc. The
instrument detects network capability and optimizes communication speed.

The computer that operates the instrument must have clear access through the network, and be
configured to communicate properly. This communication is the responsibility of your on-site
network personnel.

Full instructions are given in section 3.0.

2.6

Instrument Dimensions and Weight XDS™ OptiProbe

Analyzer dimensions are:

Width:

18.0” (457 mm)

(Leave about

6” (152mm) on the right side for access to standards door)

Height:

15.0” (381 mm)

Depth:

22.5” (572 mm) front to back

Please leave space (6 inches, 150mm) at front for fiber optics.

Leave a minimum of 3” (76mm) around the back and left side for airflow and access space. Leave at
least 6” (152mm) on the right side, and in front for fiber bundle space.

Weight:

67.2 pounds (30.6 kg)

Follow lifting instructions (on last page) when moving the instrument. Avoid injury.