Metrohm NIRS XDS Transmission OptiProbe Analyzer User Manual
Page 25
![background image](/manuals/586323/25/background.png)
▪▪▪▪▪▪▪
23
9.
Lift the release handle on the
monochromator and engage the module
“catches” to the locking togs on the
monochromator. (Module not shown to
allow a good view of handle.)
Push monochromator and module together
firmly (with handle up) then lower the
release handle.
10.
When the catches are fully engaged to the
locking togs, push the release handle down
all the way.
This automatically engages the electrical
connector and fiber optic interface, and
maintains proper alignment of the module
to the instrument.
Carefully wind the fiber bundle clockwise
around the “saddle.” The 2-meter probe
(shown) takes one wrap. The 3-meter
probe takes 2 wraps.
The final assembly is as shown.
11.
Plug the AC power cord into a grounded AC outlet. A surge protector or Uninterruptible
Power Supply (UPS) is recommended for best operation.
12.
For “Direct Connection” the gray cable from
the instrument plugs directly to the computer
network jack. (Use the UTP-crossover cable
supplied with the instrument.)
If using a network, use a non-crossover type
cable as listed in section 3.0. Plug the RJ-45
network connector into a functional network
port.
If this requires approval from a network
administrator, it should be properly approved
for hookup.