At the end of Performance Test, all measured values are compared
with acceptance criteria stored in Vision. If all results meet
acceptance criteria, the test is successful and this dialog box is
displayed.

Before clicking “Close All Reports”, the user is directed to the
tabular display.

To view the tabular display of results, place cursor over the tabular display and double-click twice.
Vision enlarges the tabular portion of the screen. Now click on the OpQual tab, near the bottom of
the screen. Noise summary results are shown.