Metrohm 920 Absorber Module User Manual
Page 85
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
920 Absorber Module
■■■■■■■■
77
Connecting the capillary for the combustion gases
1 Feeding the capillary into the Combustion Module
■
Push the loose end of the PTFE capillary 1/8 in. / 1/16 in. ID / 1.5
m (6.1803.130) (7-8) through the opening in the right-side door
of the 920 Absorber Module.
■
Guide the capillary through the round opening in the bottom of
the Combustion module into the inside.
2 Shortening the capillary
Shorten the capillary using the capillary cutter so that it can be con-
nected to the CIC T connector's outlet (23-4) easily.
3 Connecting the capillary
■
Insert the capillary into an angled FAST connector (6.7304.000).
■
Carefully attach the FAST connector to the CIC T connector's out-
let (23-4).
Connecting the capillary for the absorber solution
1 Feeding the capillary into the Combustion Module
Guide the loose end of the capillary with the yellow colored sleeve
into the inside through the round opening at the top in the left side
wall.
2 Shortening the capillary
Shorten the capillary using the capillary cutter so that it can be con-
nected to the CIC T connector's outlet (23-1) easily.
3 Connecting the capillary
■
Plug the capillary into a straight FAST connector (6.7304.010) (see
Chapter "Replacing tubing connections" in the Combustion Mod-
ule operating instructions).
■
Carefully attach the FAST connector to the CIC T connector's out-
let (23-1).