6 connecting the bottle with the absorber solution, Connecting the bottle with the absorber solution – Metrohm 920 Absorber Module User Manual
Page 60
4.4 Dosino for the absorber solution
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
920 Absorber Module
3 Connecting the column's outlet
■
Screw the PTFE capillary (6.1803.030) with the yellow colored
sleeve (fed in from the right side of the instrument) in place on
the trap column's outlet using a pressure screw (6.2744.010).
Inserting the trap column in the instrument
Insert the Metrosep A Trap 1 – 100/4.0 in the instrument as follows:
1 Fasten the Metrosep A Trap 1 – 100/4.0 in place in the front holding
bracket (4-4) of the 920 Absorber Module using the column head.
Next step
Connect the loose end of the PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030) to the Combustion Module's outlet (see Chapter 4.10.2,
page 76).
4.4.6
Connecting the bottle with the absorber solution
The bottle with the absorber solution is connected to port 2 of Dosinos 1.
Accessories
For this step you need:
■
Clear glass bottle 2 L (6.1608.070) filled with the absorber solution
■
Bottle attachment (6.1602.160)
■
FEP tubing M6 for 920, 150 cm (6.1838.020) fastened to port 2 of the
Dosino for the absorber solution.
■
Adsorber tube (6.1609.000)
■
Cotton
■
Adsorber granulate