5 connecting the metrosep a trap 1, Connecting the metrosep a trap 1 – Metrohm 920 Absorber Module User Manual
Page 57
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
920 Absorber Module
■■■■■■■■
49
4.4.5
Connecting the Metrosep A Trap 1
The Metrosep A Trap 1 – 100/4.0 (6.1014.000) trap column continues to
remove the smallest anionic contaminants in the absorber solution. The
use of trap column ensures that the analysis is not distorted by unwanted
contamination.
The Metrosep A Trap 1 – 100/4.0 is inserted between the Dosino for
absorber solution and the T connector CIC (6.7301.210) at the outlet of
the Combustion Module.
The …
… is connected to
Inlet
Dosino for the absorber solution
Outlet
Combustion tube outlet on the Combustion Module
Accessories
Connect the following accessories to the trap column:
On …
Install this accessory
Inlet
■
Approx. 20 cm of the PTFE capillary (6.1803.030) connected to
port 3 of the Dosino for the absorber solution (10-
■
Pressure screw (6.2744.010)
■
Colored sleeve, yellow
Outlet
■
PTFE capillary (6.1803.030) (10-
■
Optional: Colored sleeve, yellow
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)