beautypg.com

5 connecting the metrosep a trap 1, Connecting the metrosep a trap 1 – Metrohm 920 Absorber Module User Manual

Page 57

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

920 Absorber Module

■■■■■■■■

49

4.4.5

Connecting the Metrosep A Trap 1

The Metrosep A Trap 1 – 100/4.0 (6.1014.000) trap column continues to
remove the smallest anionic contaminants in the absorber solution. The
use of trap column ensures that the analysis is not distorted by unwanted
contamination.

The Metrosep A Trap 1 – 100/4.0 is inserted between the Dosino for
absorber solution and the T connector CIC (6.7301.210) at the outlet of
the Combustion Module.

The …

… is connected to

Inlet

Dosino for the absorber solution

Outlet

Combustion tube outlet on the Combustion Module

Accessories

Connect the following accessories to the trap column:

On …

Install this accessory

Inlet

Approx. 20 cm of the PTFE capillary (6.1803.030) connected to
port 3 of the Dosino for the absorber solution (10-

10

)

Pressure screw (6.2744.010)

Colored sleeve, yellow

Outlet

PTFE capillary (6.1803.030) (10-

9

)

Optional: Colored sleeve, yellow