beautypg.com

1 installing the drainage tubing, Installing the drainage tubing – Metrohm 920 Absorber Module User Manual

Page 79

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Installation

920 Absorber Module

■■■■■■■■

71

The leak sensor is switched to active in the software.

4.8.1

Installing the drainage tubing

Liquid that leaks in the bottle holder flows to the rear of the instrument
and is drained through an opening on the bottle holder. Drainage tubing
has to be mounted at this opening. This drainage tubing guides the leak-
ing liquid to the base tray where the leak sensor is located.

Accessories

For this step you need the following parts from the accessory kit: Vario/
Flex Basic (6.5000.000):

2 × silicone tubing (6.1816.080)

You also need scissors.

Connecting the drainage tubing

6.1816.080

6.1816.080

1 Attach one end of the silicone tubing to the drainage support on the

bottle holder.

Attach the other end of the silicone tubing to the right-side drainage
support on the base tray.

2 Attach one end of the second piece of silicone tubing to the left-side

drainage tubing connection on the base tray.