beautypg.com

5 connecting the metrosep i trap 1, Connecting the metrosep i trap 1 – Metrohm 920 Absorber Module User Manual

Page 42

background image

4.3 Dosino for ultrapure water

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

920 Absorber Module

Connect
capillary/tub-
ing at port

… to

3

T connection (see Chapter 4.4.4, page 47)

4

Combustion oven inlet (see Chapter 4.10.1, page 74)

4.3.5

Connecting the Metrosep I Trap 1

The Metrosep I Trap 1 – 100/4.0 trap column (6.1014.200) removes dis-
ruptive interfering anions and cations from the ultrapure water used for
transferring the absorption solution from the absorber tube to the 6-port
valve. The use of trap column ensures that the analysis is not distorted by
unwanted contamination.

The Metrosep I Trap 1 – 100/4.0 is used between the Dosino for ultrapure
water and the 6-port valve.

The …

… is connected to

Inlet

Dosino for ultrapure water

Outlet

6-port valve

Accessories

The following accessories have to be connected to the trap column:

On …

Install this accessory

Inlet

30 cm of the PTFE capillary (6.1803.030) (8-

16

)

Pressure screw (6.2744.010)

Outlet

40 cm of the PTFE capillary (6.1803.030) (8-

17

)

Pressure screw (6.2744.010)

Connecting the Metrosep I Trap 1 – 100/4.0

NOTE

The trap column is sealed with threaded stoppers at both ends.

Remove these threaded stoppers before connecting the capillaries.