Metrohm 920 Absorber Module User Manual
Page 27
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
920 Absorber Module
■■■■■■■■
19
9
PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Connection between the Metrosep A Trap 1
- 100/4.0 trap column (6.1014.000) and the
Combustion Module.
10 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Approx. 20 cm. Connection between the
Dosino with the absorber solution and the
Metrosep A Trap 1 - 100/4.0 trap column
(6.1014.000).
11 FEP tubing M6 for 920, 150 cm
(6.1838.020)
Aspiration tubing for the absorber solution.
Connection between the Dosino with the
absorber solution and the bottle with the
absorber solution.
12 FEP tubing M6 for 920, 9 cm
(6.1838.040)
Connection between the Dosino with the
absorber solution and the T connector.
13 FEP tubing M6 for 920, 9 cm
(6.1838.040)
Connection between the Dosino with ultra-
pure water and the T connector.
14 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Connection between the Dosino with ultra-
pure water and the Combustion Module.
15 FEP tubing M6 for 920, 150 cm
(6.1838.020)
Aspiration tubing for ultrapure water. Con-
nection between the Dosino with the water
bottle.
16 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Approx. 30 cm. Connection between the
Dosino with ultrapure water and the Metro-
sep I Trap 1 - 100/4.0 trap column
(6.1014.200).
17 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Approx. 40 cm. Connection between the
Metrosep I Trap 1 - 100/4.0 trap column
(6.1014.200) and the 6-port valve.
18 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Connection between the 6-port valve and
the 930 Compact IC Flex.
Colored sleeves
Different liquids move throughout the 920 Absorber Module. In order to
provide you an overview of the different flow paths, we recommend using
the provided colored sleeves as follows to code the capillaries that route
the distinct liquids:
Table 1
Recommended color codes for the flow paths
Liquid
Colored sleeve
Ultrapure water
blue
Absorber solution
yellow
Standard solution
black
Check standard
red
Waste
white