2.2 System description
■■■■■■■■■■■■■■■■■■■■■■
8
■■■■■■■■
920 Absorber Module
2.2
System description
Depending on the application, the Combustion IC system consists of the
following instruments:
Type of sample
Instruments
Manufacturer
solid/liquid samples only
■
Combustion Module with Auto Boat Drive (ABD)
(2.136.0700)
■
optional: sampler (MMS 5000)
Analytik Jena
■
920 Absorber Module (2.920.0010)
■
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
Metrohm
gaseous samples only
■
Combustion Module with LPG/GSS Module (2.136.0730)
Analytik Jena
■
920 Absorber Module (2.920.0010)
■
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
Metrohm
solid, liquid and gaseous
samples
■
Combustion Module with Auto Boat Drive (ABD)
(2.136.0700)
■
LPG/GSS Module (2.136.0720)
Analytik Jena
■
920 Absorber Module (2.920.0010)
■
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
Metrohm
Figure 2
Overview - CIC system
The sample is combusted in the Combustion Module.
The gases generated during combustion are dissolved into an absorber
solution in the 920 Absorber Module and are optimized for analysis using
suitable sample preparation techniques.
In the end, the absorption solution is preconcentrated in an ion chromato-
graph and analyzed.
The entire process is controlled and monitored using MagIC Net software.
The analysis data is saved in the software's database.