1 connecting the ports, Connecting the ports – Metrohm 920 Absorber Module User Manual
Page 63
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
920 Absorber Module
■■■■■■■■
55
4
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Approx. 80 cm. Connection between the
10-port valve and the bottle with the check
standard.
5
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Approx. 60 cm. Connection between the
10-port valve and the absorber tube.
6
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Approx. 5 cm. For aspirating air segments.
7
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Connection between the 10-port valve and
the waste container.
4.5.1
Connecting the ports
The transfer tubing and capillaries for all of the liquids being used are con-
nected to the 10-port valve.
Connectors
The ports of the 10-port valve are used as follows:
Port …
… is used for …
… is connected to
Central
Aspirating and discharging differ-
ent solutions
Transfer tubing
1
Outlet for sample, standard solu-
tion, check standard
6-port valve
2
Inlet for standard solution
Bottle with standard solution
3
Inlet for check standard
Bottle with check standard
4-7
Not used
–
8
Inlet for sample
Absorber tube
9
Aspirating air segments
–
10
Outlet for waste
Waste container
Accessories
The following accessories have to be connected to the 10-port valve:
Port
Install this accessory
Central
■
Adapter UNF 10/32 outer / M6 inner (6.2744.200)
■
Transfer tubing (6.1562.130)
1
■
PEEK capillary, 0.5 mm ID / 15 cm (6.1831.040) (13-
■
Pressure screw (6.2744.090)
2
■
PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 80 cm
(13-
■
Pressure screw (6.2744.090)
3
■
PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 80 cm
(13-
■
Pressure screw (6.2744.090)
4-7
Threaded stopper (6.2744.220)