Figure 14, Connecting the 10-port valve, Figure 14 connecting the 10-port valve – Metrohm 920 Absorber Module User Manual
Page 64: 1 connecting the central port, Screw the adapter in place in the central port, 2 connecting port 1
4.5 10-port valve
■■■■■■■■■■■■■■■■■■■■■■
56
■■■■■■■■
920 Absorber Module
Port
Install this accessory
8
■
PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 60 cm
(13-
■
Pressure screw (6.2744.090)
9
■
PTFE capillary, 0.75 mm ID / 3 m (6.1803.160), approx. 5 cm
(13-
■
Pressure screw (6.2744.090)
10
■
PTFE capillary, 0.75 mm ID / 3 m (6.1803.160) (13-
■
Pressure screw (6.2744.090)
Connecting the 10-port valve
Connect the ports of the 10-port valve according to the following figure:
NOTE
We recommend connecting the central port first.
6.2744.090
6.1831.040
6.2744.090
6.1803.160
80 cm
6.1803.160
80 cm
6.2744.090
6.27440220
6.2744.220
6.2744.220
6.2744.220
6.2744.090
6.1803.160
60 cm
6.2744.090
6.1803.160
5 cm
6.2744.090
6.1803.160
3 m
6.2744.200
6.1462.130
1
2
3
4
5
6
7
8
9
10
Figure 14
Connecting the 10-port valve
1 Connecting the central port
■
Screw the adapter in place in the central port.
■
Screw one end of the transfer tubing in place on the adapter.
2 Connecting port 1
■
Screw the PEEK capillary to port 1 using a long pressure screw.