4.3 Dosino for ultrapure water

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

920 Absorber Module

Cut off a piece from the PTFE capillary (6.1803.030) that is
approx. 40 cm long using the capillary cutter (6.2621.080) and
tighten it onto the outlet of the trap column using a pressure
screw.

Inserting the trap column in the instrument

Insert the Metrosep I Trap 1 – 100/4.0 in the instrument as follows:

1 Fasten the Metrosep I Trap 1 – 100/4.0 in place in the rear holding

bracket (4-4) of the 920 Absorber Module using the column head.

Next step

Tighten the loose end of the PTFE capillary at the outlet of the trap col-
umn to port 5 of the 6-port valve (see Chapter 4.6, page 61).

4.3.6

Connecting the ultrapure water bottle

The ultrapure water bottle has to be connected to port 2 of the Dosino
for ultrapure water.

Accessories

For this step you need:

Clear glass bottle 2 L (6.1608.070) filled with ultrapure water.

Bottle attachment (6.1602.160)

FEP tubing M6 to 920 150 cm (6.1838.020) (8-15).

Adsorber tube (6.1609.000)

Cotton

Adsorber granulate