beautypg.com

Metrohm 920 Absorber Module User Manual

Page 52

background image

4.4 Dosino for the absorber solution

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

920 Absorber Module

FEP tubing (6.1838.020) is threaded in through the thread adapter
(6.1618.020) and the Dosino holder (6.2057.230) and is connected to
port 2.

1 Tighten the Dosino holder (6.2057.230) with the thread adapter

(6.1618.020) on the bottom of the dosing unit.

Next step

Connect the ports 1 and 3.

Connecting port 1 and port 3

Accessories

The following accessories have to be connected to these two ports:

Port

Install this accessory

Port 1

FEP tubing M6 for 920, 9 cm (6.1805.040) (10-

12

)

Port 3

PTFE capillary (6.1803.030), approx. 20 cm (10-

10

)

Pressure screw (6.2744.350)

Coupling (6.2744.080)