beautypg.com

Care and use manual – Waters IC-Pak C Method Development Column User Manual

Page 2

background image

[ Care and Use ManUal ]

Waters IC-Pak C M/D Column

2

II. InstALLInG tHe CoLuMn

a. Preparing the Ion-Exchange System

To prepare the system, you need two 3/8-inch wrenches and a zero
dead-volume union. Before installing the column in the flow path:

1. Directly connect the injector to the detector by replacing the old column

with a zero dead-volume union.

2. Flush the lines and the detector of microparticulates and previous sol-

vents. Use fresh eluent at a flow rate of 1.0 mL/min. Flush the injector
loop if applicable.

3. Remove the union.

4. Install the column between the injector and detector (see Section II. b).

b. Installing the Column

Remove the end plugs from the column and save them for use when
you store the column. To install the column:

1. Thread the inlet and outlet fittings into the column until finger-tight.

An arrow on the label indicates the direction of flow.

2. Use the wrenches to tighten the fittings 1 /4 to 1 /2 turn. Do not

overtighten. Overtightening damages the connection.

3. Prepare a new tubing-ferrule connection (Figure 1) when connecting a

new column or when removing a damaged compression screw or worn
ferrule.

Note: The tubing distance beyond the ferrule may differ for each
column type. Resize the tubing to the correct distance by replacing
the ferrule.

Preparing a New Connection

To prepare a new tubing-ferrule connection:
1. Scribe the circumference of the tubing at the desired break using either

a file with a cutting edge or a tube cutter.

2. Grasp the tubing on both sides of the scribe mark with cloth-covered

pliers (to prevent marring the tube surface), and gently work the tube
back and forth until it separates.

Note: Ensure that tubing end is straight, open, and free of burrs.

3. Slide the compression fitting, followed by the ferrule (large end of the

taper first) over the tube.

Note: Properly bottom the tubing in the fitting seat. Otherwise, dead
volume could result in sample bond spreading.

Figure 1: Ferrule and Compression Screw Assembly

c. Equilibrating the Column

1. Precondition new columns by flushing with 100% acetonitrile for 30

minutes at 0.5 mL/min.

2. Flush the column with mobile phase for 5 minutes at 0.5 mL/min,

and them for 30 minutes at 1.0 mL/min before performing the first
separation.

III. PrePArInG MoBILe PHAse And sAMPLes

a. Preparing the Mobile Phase

Chemical requirements

The IC-Pak C M/D uses 0.1 mM EDTA/3.0 mM HNO

3

as a mobile

phase. Store mobile phase at room temperature for up to one month.

ACS Reagent grade or equivalent quality free-acid EDTA

Quality nitric acid

Quality 18 megohm water

Plasticware to prepare and store all mobile phases, standards, and
samples

Compression Screw or Nut

Ferrule

0.009-inch ID Tubing

(0.23 mm)