beautypg.com

Care and use manual, Ii. connecting the column to the hplc instrument, Uv @ 254 nm – Waters PAH Columns User Manual

Page 2

background image

[ Care and Use ManUal ]

Name of Document

2

Column:

Waters PAH Column 5 μm 4.6 x 250 mm @ 27 °C

Peaks:

System:

Waters Alliance System with 2996 Photodiode Array Detector

Eluent A:

Water

Eluent B:

Acetonitrile

Gradient:

60% B to 100% B using curve 9 in 12 minutes,

hold 11 minutes, back to initial conditions

Flow Rate:

1.2 mL/min

Injection:

20 μL

Sample:

EPA-610 mixture

UV @ 254 nm

Figure 1: PAH Analysis using Waters PAH Columns

Column:

Waters PAH Column 5 μm 4.6 x 250 mm @ 27 °C

Peaks:

Eluent A:

Water

Eluent B:

Acetonitrile

Gradient:

60% B to 100% B using curve 9 in 12 minutes

hold 11 minutes, back to initial conditions

Flow Rate:

1.2 mL/min

Injection:

20 μL

Sample:

EPA-610 mixture plus two compounds*

UV @ 254 nm

Figure 2: PAH Analysis According to Florida Administrative Code 17,700

II. CONNECTING THE COLUMN TO THE HPLC INSTRUMENT

Handle the column with care. Do not drop or hit column on a hard surface
as it may disturb the bed and affect its performance.

1. Correct connection of 1/16 inch outer diameter stainless steel tubing

leading to and from the column is essential for high-quality chromato-

graphic results.

9. Benzo(a)anthracene - 2 ppm
10. Chrysene - 2 ppm
11. Benzo(b)fluoranthene - 4 ppm
12. Benzo(k)fluoranthene - 2 ppm
13. Benzo(a)pyrene - 2 ppm
14. Dibenzo(a,h)anthracene - 4 ppm
15. Benzo(g,h,I)perylene - 4 ppm
16. Indeno(1,2,3-cd)pyrene - 2 ppm

10. Pyrene - 2 ppm
11. Benzo(a)anthracene - 2 ppm
12. Chrysene - 4 ppm
13. Benzo(b)fluoranthene - 4 ppm
14. Benzo(k)fluoranthene - 2 ppm
15. Benzo(a)pyrene - 2 ppm
16. Dibenzo(a,h)anthracene - 4 ppm
17. Benzo(g,h,I)perylene - 4 ppm
18. Indeno(1,2,3-cd)pyrene - 2 ppm

2. When using standard stainless steel compression screw fittings, it is

important to ensure proper fit of the 1/16 inch outer diameter stainless
steel tubing. When tightening or loosening the compression screw, place
the 5/16 inch wrench on the compression screw and the other 3/8 inch
wrench on the hex head of the column endfitting. Note: If one of the
wrenches is placed on the column flat during this process, the endfitting
will be loosened and leak.

Minutes

24

6

8

10

12

14

16

18

20

22

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

UV @ 254 nm

9

11

13

15

17

19

21

23

25

1

2

3 *

4 *

5

6

7

8

9 10

11

12

13

14

15

16

17

18

Minutes

5

7

UV @ 254 nm

1. Naphthalene - 20 ppm
2. Acenaphthylene - 40 ppm
3. Acenaphthene - 20 ppm
4. Fluorene - 4 ppm
5. Phenanthrene - 2 ppm
6. Anthracene - 2 ppm
7. Fluoranthene - 4 ppm
8. Pyrene - 2 ppm

1. Naphthalene - 20 ppm
2. Acenaphthylene - 40 ppm
3*. 1-methyl naphthalene - 25 ppm
4*. 2-methyl naphthalene - 25 ppm
5. Acenaphthene - 20 ppm
6. Fluorene - 4 ppm
7. Phenanthrene - 2 ppm
8. Anthracene - 2 ppm
9. Fluoranthene - 4 ppm