beautypg.com

Waters XBridge Peptide BEH C18 130A and 300A Columns User Manual

Xbridge peptide beh c, Care and use manual

background image

CONT ENTS
I. GET TING START ED

a. Column Installation

b. Column Equilibration

c. Initial Column Efficiency Determination

II. COLUMN USE

a. Sample Preparation

b. Operating pH Limits

c. Solvents

d. Pressure

e. Temperature

III. SCALING UP/DOWN ISOC RATIC MET HODS

IV. T ROUBLESHOOTING

V. COLUMN CLEANING, REGENERATING

AND STORAGE

a. Cleaning and Regeneration

b. Storage


VI. CONNECTING T HE COLUMN TO T HE HPLC

a. Column Connectors and System Tubing Considerations

b. Measuring System Bandspreading Volume

c. Measuring Gradient Delay Volume (or Dwell Volume)

VII. ADDITIONAL INFORMATION

a. Use of Narrow-Bore (3.0 mm i.d.)

b. Impact of Bandspreading Volume on 2.1 mm i.d.

Column Performance

c. Non-Optimized vs. Optimized LC-MS/MS System:

System Modification Recommendations

Thank you for choosing a Waters XBridge

®

Peptide BEH C

18

, 130Å

or 300Å Column. The XBridge Peptide BEH C

18

, 130Å and 300Å

packing materials were designed to provide excellent peak shape,
high efficiency and excellent stability. The XBridge Peptide BEH C

18

,

130Å and 300Å packing materials are manufactured in a cGMP,
ISO 9002 certified planted using ultra pure reagent. Each batch
of XBridge Peptide BEH C

18

Column material has been qualified

with a peptide separation and the results are held to narrow
specification ranges to assure excellent, reproducible performance
for peptide separations. Every column is tested and a Performance
Test Chromatogram along with a Certification of Acceptance are
provided with each column.

XBridge Peptide BEH C

18

, 130Å and 300Å Columns

[ CARE AND USE MANUAL ]