beautypg.com

Care and use manual – Waters Peptide Separation Technology ACQUITY UPLC BEH130 and BEH300 User Manual

Page 6

background image

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com

[ CARE AND USE MANUAL ]

©2014 Waters Corporation. Waters, T he Science of W hat’s
Possible, Oasis, Sep-Pak, Empower, MassLynx, and XBridge
are registered trademarks of Waters Corporation. eCord is a
trademarks of Waters Corporation. All other trademarks are
property of their respective owners.

February 2014 715001465 Rev C VW-PDF

T1

T8

T15

T5

T19

T12-T13

T12

T4

T9-T10

T10

T14

T13-T14

T19C

Minutes

6.00

9.00

12.00

15.00

18.00

21.00

24.00

*Chromatographic Conditions:
Column:

2.1 mm x 100 mm,

Sample:

Tryptic digest of cytochrome c;

Flow rate:

0.20 mL/min

Mobile phase A: 0.045% TFA in water;
Mobile phase B: 0.045% TFA in acetonitrile;
Gradient:

from 0 to 15% B in 6 min, from 15 to

36% B in 20 min;

Temperature: 35 ˚C; Detection: UV 214 nm

VI. REP RESENTATIV E T EST CHROMATOGRAPH AND CONDITIONS FOR SEPARATION/

OF P ROT EIN DIGEST

Peak Identification

T1

N-AcGDVEK

T8

TGPNLHGLFGR

T13-T14

KYIPGTK

T15

MIFAGIK

T14

YIPGTK

T5

CAQCHTVEK (heme attached)

T4

IFVQK

T19

EDLIAYLK

T9-T10

KTGQAPGFSYTDANK T12-T13

GITWGEETLMEYLENPKK

T10

TGQAPGFSYTDANK

T12

GITWGEETLMEYLENPK

T19C

EDLIAY

This manual is related to the following products: