Care and use manual – Waters Peptide Separation Technology ACQUITY UPLC BEH130 and BEH300 User Manual
Page 6
Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com
[ CARE AND USE MANUAL ]
©2014 Waters Corporation. Waters, T he Science of W hat’s
Possible, Oasis, Sep-Pak, Empower, MassLynx, and XBridge
are registered trademarks of Waters Corporation. eCord is a
trademarks of Waters Corporation. All other trademarks are
property of their respective owners.
February 2014 715001465 Rev C VW-PDF
T1
T8
T15
T5
T19
T12-T13
T12
T4
T9-T10
T10
T14
T13-T14
T19C
Minutes
6.00
9.00
12.00
15.00
18.00
21.00
24.00
*Chromatographic Conditions:
Column:
2.1 mm x 100 mm,
Sample:
Tryptic digest of cytochrome c;
Flow rate:
0.20 mL/min
Mobile phase A: 0.045% TFA in water;
Mobile phase B: 0.045% TFA in acetonitrile;
Gradient:
from 0 to 15% B in 6 min, from 15 to
36% B in 20 min;
Temperature: 35 ˚C; Detection: UV 214 nm
VI. REP RESENTATIV E T EST CHROMATOGRAPH AND CONDITIONS FOR SEPARATION/
OF P ROT EIN DIGEST
Peak Identification
T1
N-AcGDVEK
T8
TGPNLHGLFGR
T13-T14
KYIPGTK
T15
MIFAGIK
T14
YIPGTK
T5
CAQCHTVEK (heme attached)
T4
IFVQK
T19
EDLIAYLK
T9-T10
KTGQAPGFSYTDANK T12-T13
GITWGEETLMEYLENPKK
T10
TGQAPGFSYTDANK
T12
GITWGEETLMEYLENPK
T19C
EDLIAY