Care and use manual – Waters Peptide Separation Technology ACQUITY UPLC BEH130 and BEH300 User Manual
Page 2
2
[ CARE AND USE MANUAL ]
ACQUITY UPLC Peptide BEH C
18
, 130Å and 300Å Columns
II. GET TING START ED
Each Peptide Separation Technology ACQUITY UPLC Peptide BEH
C
18
, 130Å and 300Å Columns comes with Certificate of Analysis
and a Performance Test Chromatogram embedded within the eCord
intelligent chip. The Certificate of Analysis is specific to each batch
of packing material contained in the ACQUITY UPLC Peptide BEH
C
18
, 130Å and 300Å Columns and includes the gel batch number,
analysis of unbonded particles, analysis of bonded particles, and
chromatographic results and conditions. The Performance Test
Chromatogram is specific to each individual column and contains
such information as: gel batch number, column serial number, USP
plate count, USP tailing factor, capacity factor, and chromatographic
conditions. These data should be stored for future reference.
a. Column Connectors
The ACQUITY UPLC System utilizes tubing and gold plated
compression screws which have been designed to meet stringent
tolerance levels and to minimize extra column volumes.
Optimized column inlet tubing (Part Number 430001084) is
supplied with the ACQUITY UPLC System. The inject valve end of
the tubing is clearly marked with a blue shrink tube marker. Insert
the opposite end of the tubing into the ACQUITY UPLC Column and
tighten the compression fitting using two 5/16-inch wrenches.
For information on the correct column outlet tubing, please refer
to the relevant detector section in the ACQUITY UPLC System
Operator’s Guide (Part Number 71500082502).
b. Column Installation
1. Purge the pumping system of any buffer-containing mobile phases
and connect the inlet end of the column to the injector outlet.
2. Flush column with 100% organic mobile phase (methanol or
acetonitrile) by setting the pump flow rate to 0.1 mL/min and
increase the flow rate to 0.5 mL/min over 5 minutes.
3. When the mobile phase is flowing freely from the column
outlet, stop the flow and attach the column outlet to the
detector. This prevents entry of air into the detection system
and gives more rapid baseline equilibration.
4. Gradually increase the flow rate as described in step 2.
5. Once a steady backpressure and baseline have been achieved,
proceed to the next section.
Note: If mobile phase additives are present in low concentrations (e.g.,
ion-pairing reagents), 100 to 200 column volumes may be required
for complete equilibration. In addition, mobile phases that contain
formate (e.g., ammonium formate, formic acid, etc.) may also require
longer initial column equilibration times.
c. Column Equilibration
ACQUITY UPLC BEH130 and BEH300 Columns are shipped in 100%
acetonitrile. It is important to ensure mobile phase compatibility
before changing to a different mobile phase system. Equilibrate the
column with a minimum of 10 column volumes of the mobile phase to
be used (refer to Table 1 for a list of column volumes)
Table 1. Empty Column Volumes in mL (Multiply by 10 for
Flush Solvent Volumes)
Column Length (mm)
Internal Diameter 2.1 mm
50
0.2 mL
100
0.4 mL
150
0.5 mL
To avoid precipitating mobile phase buffers on your column or
in your system, flush the column with five column volumes of a
water/organic solvent mixture, using the same or lower solvent
content as in the desired buffered mobile phase. (For example,
flush the column and system with 60% methanol in water prior to
introducing 60% methanol/40% buffer mobile phase.)
d. eCord Installation
The eCord button should be attached to the side of the column
heater module. The eCord button is magnetized and does not
require specific orientation.
e. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it.
Waters recommends using a suitable solute mixture, as found
in the “Performance Test Chromatogram”, to analyze the
column upon receipt.
2. Determine the number of theoretical plates (N) and use this
value for periodic comparisons.
3. Repeat the test at predetermined intervals to track column
performance over time. Slight variations may be obtained
on two different UPLC systems due to the quality of the
connections, operating environment, system electronics,
reagent quality, column condition and operator technique.