[ Care and Use ManUal ]

Oasis Prospekt-2/Symbiosis SPE Cartridges

2. If necessary, clarify samples by centrifugation at 8,000 x g

for 20 minutes.

Note: The HPLC or LC/MS/MS method is normally developed before
the SPE method. If possible, select a mobile phase that can elute the
analyte(s) from the Oasis SPE cartridge and also gives the required
chromatography on the analytical column.

b. Prepare the Prospekt-2/Symbiosis System According to the
Operator’s Manual

1. Place three solvents in the high pressure dispenser (HPD) for the

Prospekt-2/Symbiosis system: a) 100% methanol, b) water, and
c) 5% methanol in water.

2. Place the Oasis cartridge tray in the Automatic Cartridge

Exchange module.

3. Place prepared samples in the autosampler.

4. Set up the HPLC system with the mobile phase and analytical

column.

Note: If you encounter problems with your Prospekt-2/Symbiosis
systems, contact your Spark Holland representative.

III. strategIes for method oPtImIzatIon on oasIs
methodoLogIes

(Please refer to www.waters.com/oasis)

The following tips can be employed to optimize on-line SPE methods
in general.

1. Increase the flow rates from solvation, equilibration, and wash to

shorten the process time.

2. Do not exceed the recommended backpressure of 300 bar (4,500 psi).

3. To improve extraction efficiency, reduce the flow rate during the

loading step (depending on the analyte).

4. Use solvent mixes having greater or lesser polarity to produce a

cleaner extract (for example, 60/40:IPA/acetonitrile).

5. For very polar analytes, use 100% pH-modified water for wash

and elution solution.

6. For very non-polar analytes, use stronger solvents, such as

70/30:acetonitrile/methonol.

Note: The achiever better cleanup, optimized method(s) can be used.

a. On-line SPE Generic Starting Methods for Oasis Prospekt-2/
Symbiosis Cartridges

These methods serve as a generic starting point for developing Oasis
Prospekt-2/Symbiosis SPE protocol; optimization may be required.
(For further information, please refer to www.waters.com/oasis)

Oasis HLB Cartridges

Solvation: 100% methanol

Flow rate: 5 mL/min

Volume: 1,000 µL

Equilibrate: 100% water

Flow rate: 5 mL/min

Volume: 1,000 µL

Load: 100% water

Flow rate: 1.0 mL/min

Inject: 20–100 µL of sample

Volume: 500 µL

Wash 1: 5% methanol in water

Flow rate: 2 mL/min

Volume: 1,000–2,000 µL

Wash 2: 100% water

Flow rate: 5 mL/min

Volume: 1,000–2,000 µL

Elute: HPLC mobile phase

Flow rate: 0.5 mL/min