b. Operating pH Limits

The recommended operating pH range for ACQUITY UPLC BEH300, C

4

, 1.7 µm columns is 1 to 12. A listing of commonly used buffers and

additives is given in Table 2. Additionally, the column lifetime will vary depending upon the operating temperature as well as the type and
concentration of buffer used.

Table 2: Buffer Recommendations for Using ACQUITY UPLC BEH300, C

4

columns from pH 1 to 12

Additive/Buffer

pKa

Buffer range

(±1 pH unit)

Volatility

Used for

Mass Spec

Comments

TFA

0.3

Volatile

Yes

Ion pair additive, can suppress MS signal, used in the 0.02-0.1% range.

Acetic Acid

4.76

Volatile

Yes

Maximum buffering obtained when used with ammonium acetate salt. Used in 0.1-1.0% range.

Formic Acid

3.75

Volatile

Yes

Maximum buffering obtained when used with ammonium formate salt. Used in 0.1-1.0% range.

Acetate

(NH

4

CH

2

COOH)

4.76

3.76 – 5.76

Volatile

Yes

Used in the 1-10 mM range. Note that sodium or potassium salts are not volatile.

Formate

(NH

4

COOH)

3.75

2.75 – 4.75

Volatile

Yes

Used in the 1-10 mM range. Note that sodium or potassium salts are not volatile.

Phosphate 1

2.15

1.15 – 3.15

Non-

volatile

No

Traditional low pH buffer, good UV transparency.

Phosphate 2

7.2

6.20 – 8.20

Non-

volatile

No

Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.

Phosphate 3

12.3

11.3 - 13.3

Non-

volatile

No

Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime.

4-Methylm-

orpholine

~8.4

7.4 – 9.4

Volatile

Yes

Generally used at 10 mM or less.

Ammonia

(NH

4

OH)

9.2

8.2 – 10.2

Volatile

Yes

Keep concentration below 10 mM and temperatures below 30 ˚C.

Ammonium
Bicarbonate

10.3

(HCO3

-

)

9.2

(NH4

+

)

8.2 – 11.3

Volatile

Yes

Used in the 5-10 mM range (for MS work keep source >150 ˚C). Adjust pH with ammonium
hydroxide or acetic acid. Good buffering capacity at pH 10

Note: use ammonium bicarbonate (NH

4

HCO

3

), not ammonium carbonate ((NH

4

)

2

CO3)

Ammonium

(Acetate)

9.2

8.2 – 10.2

Volatile

Yes

Used in the 1-10 mM range.

Ammonium

(Formate)

9.2

8.2 – 10.2

Volatile

Yes

Used in the 1-10 mM range.

Borate

9.2

8.2 – 10.2

Non-

Volatile

No

Reduce temperature/ concentration and use a guard column to maximize lifetime.

CAPSO

9.7

8.7 – 10.7

Non-

Volatile

No

Zwitterionic buffer, compatible with acetonitrile, used in the 1-10 mM range. Low odor.

Glycine

2.4, 9.8

8.8 – 10.8

Non-

Volatile

No

Zwitterionic buffer, can give longer lifetimes than borate buffer.

1-Methyl-
piperidine

10.2

9.3 – 11.3

Volatile

Yes

Used in the 1-10 mM range.

CAPS

10.4

9.5 – 11.5

Non-

Volatile

No

Zwitterionic buffer, compatible with acetonitrile, used in the 1-10 mM range. Low odor.

Triethylamine

(as acetate salt)

10.7

9.7 – 11.7

Volatile

Yes

Used in the 0.1-1.0% range. Volatile only when titrated with acetic acid (not hydrochloric or phos-
phoric).Used as ion-pair for DNA analysis at pH 7-9.

Pyrrolidine

11.3

10.3 – 12.3

Volatile

Yes

Mild buffer, gives long lifetime.