Care and use manual, Short elution gradient 2.1 x 50 mm, Iv. typical chromatographic conditions – Waters Oasis HLB Direct Connect HP 2.1 x 30 mm Column for Automated Sample Preparation with Waters UPLC with Online SPE Technology User Manual
Page 2

[ Care and Use ManUal ]
Oasis HLB Direct Connect HP 2.1 x 30 mm Column
2
IV. TyPICaL CHROmaTOgRaPHIC CONDITIONS
The analytical process begins with loading a sample on to the extraction 
column via the SPE injector port (valve A). The loading pump (QSM) 
using 100% aqueous mobile phase (with or without additives) at 
a flow rate of 2 mL/min moves the sample from the injection loop 
on to extraction column A. During the loading step, analytes of 
interest will be trapped on the extraction column, while un-retained 
interferences will be discarded to waste. After completing the loading 
step, the extraction column is washed to remove mildly retained inter-
ferences. After the wash step, the trapped analytes on extraction column 
A are eluted with a gradient from the ACQUITY BSM on to a UPLC ana-
lytical column for peak re-focusing. During the elution step of extraction 
column A, extraction column B is regenerated then re-equilibrated 
(using the QSM) to original conditions for the next analysis. The events 
on each of the parallel extraction columns are shown in Figure 2.
Figure 2: Timeline showing the events occurring on each of the 
extraction columns during an online SPE UPLC analysis
Typical mobile phase conditions to use for the load, wash and elu-
tion of the Oasis HLB Direct Connect HP 2.1 x 30 mm column will 
depend on the analytes of interest. The pH of the mobile phase will 
affect the retention behavior of analytes.
For a typical application consisting primarily of basic compounds, 
such as the analysis of pharmaceuticals in water, a loading condi-
tion using a high pH aqueous mobile phase would usually provide 
good retention of the analytes of interest. An example of this type 
of loading mobile phase is 2% ammonium hydroxide in water, 
pH 10. To wash mildly retained interferences from the extraction 
column without eluting the compounds of interest the percentage 
of organic in the mobile phase can be increased without changing 
the pH. For the example of basic pharmaceuticals in water, the wash 
could be set at 20% methanol with 2% ammonium hydroxide.
Figure 2: Timeline showing the events occurring on each of the extraction columns during an online SPE UPLC analysis
6
Time
Time
Time
5
4
3
2
1
0
-2
-4
-6
Acquisition
Start
Loading
SPE
wash
Flush
Flush
Re-equilibration
Recon.
5%
95%
SPE A
SPE B
Elution
Re-equilibration
6
5
4
4.04
4.36
3
2
1
6
5
4
3
2
1
0
Short Elution Gradient
2.1 x 50 mm
