beautypg.com

Care and use manual – Waters XBridge XP 2.5 µm Columns User Manual

Page 13

background image

XBridge XP 2.5 µm Columns

13

[ CARE AND USE MANUAL ]

c. Add triethylamine (TEA) or ammonium hydroxide
(NH

4

OH) modifiers to both mobile-phase (e.g., A2,

B2, etc.) reservoirs.

d. For LC/ELSD separations of mono- and/or disaccharides, typical

isocratic conditions include:

i. 75% (acetonitrile) with 0.2% TEA, 35 °C,

0.13 mL/min, 2.1 x 50 mm BEH Amide column;

ii. 77% acetone with 0.05% TEA, 85 °C,

0.15 mL/min, 2.1 x 50 mm BEH Amide column;

iii. 75% acetonitrile with 0.2% TEA, 35 °C,

0.2mL/min, 2.1 x 100 mm BEH Amide column.

e. For LC/ELSD separations of more complex sugar mixtures (e.g.,

polysaccharides), typical gradient conditions include (add TEA

modifier to both mobile-phases A and B):

i. Gradient going from 80% - 50% acetonitrile

with 0.2% TEA in 10 minutes, 35 °C, 0.13 mL/min,

2.1 x 100 mm BEH Amide column;

ii. 80% - 55% acetone with 0.05% TEA in

10 minutes, 85 °C, 0.15 mL/min, 2.1 x 100 mm BEH

Amide column.

f.

For LC/MS separations of mono- and disaccharides, typical

isocratic conditions include:

i. 75% acetonitrile with 0.1% ammonium

hydroxide, 35 °C, 0.13 mL/min, 2.1 x 50 mm

BEH Amide column.

g. For LC/MS separations of more complex sugar mixtures (e.g.,

polysaccharides), typical gradient conditions include (add NH

4

OH

modifier to both mobile-phases A and B):

i. Gradient going from 75% - 45% acetonitrile

with 0.1% ammonium hydroxide in 10 minutes,

35 °C, 0.2 mL/min, 2.1 x 100 mm BEH Amide

column.

5. More complex sample mixtures may require the use of gradient

conditions and/or longer column lengths.

6. If acetone is used in one or more mobile-phases, do not use acetone

as a sample diluent or needle wash solvent. Refer to injection solvents
section for sample diluent recommendations and miscellaneous tip
(#3) for needle wash solvent/purge solvent recommendations.

7. Typical sample preparation suggestions for samples that contain

sugars /saccharides /carbohydrates:

a. Liquid Samples

i. Dilute with 50/50 acetonitrile/water

ii. Filter using 0.45 μm or 0.22 μm syringe filter

(if necessary).

b. Solid Samples

i. Weigh out sample (~3 g) into 50 mL centrifuge tube

ii. Add 25 mL of 50/50 acetonitrile/water and

homogenize (mechanically)

iii. Centrifuge at 3200 rpm for 30 minutes

iv. Collect supernatant and filter using 0.45 μm or

0.22 μm syringe filter (if necessary).

c. Depending on sample and/or analyte concentrations, additional

sample dilutions may be necessary.

d. More complex samples and/or lower analyte concentrations may

require additional sample preparation steps and/or procedures

such as solid-phase extraction (SPE).

e. Consider VanGuard BEH Amide Pre-Columns for UPLC Column
protection.