typical digest using xbridge beh 130 c

18

column

typical digest using an acQuIty uPlc beh300 c

18

column

AU

4

8

12

16

20

24

min

0.00

0.10

0.20

0.30

0.40

0.50

T1

T13T14 T14

T4

T9T10

T10

T19C

T15

T8

T5

T19

T12T13

T12

typical digest using an acQuIty uPlc beh130 c

18

column

AU

4

8

12

16

20

24

min

0.00

0.08

0.16

0.24

T1

T13T14

T14

T4

T9T10

T10

T19C

T8

T15

T5

T19

T12T13

T12

Column:

ACQUITY UPLC BEH130 C

18

, 2.1 x 100 mm, 1.7 μm

Part number:

186003555

Mobile phase A: 0.045% TFA in water
Mobile phase B: 0.045% TFA in acetonitrile
Temp.:

35 °C

Wavelength:

214 nm

Injection vol.:

5 µL

Run time:

61 minutes

Gradient:

Time Flow %A1 %B1 Curve

0.2 100 0

6 0.2 85 15 6

26 0.2 64 36 6

40 0.2 40 60 11

50 0.2 100 0 11

70 0 100 0 11

typical digest using xbridge beh 300 c

18

column

6

10

14

18

22

26

30

34

38 min

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

0.45

AU

T1

T13-

T14 and T14

T4

T9

-T

10

T10

T19C

T15

T8

T5

T19

T12-

T13

T12

Column:

XBridge BEH300 C

18

, 4.6 x 150 mm, 3.5 μm

Part number:

186003613

Mobile phase A: 0.055% TFA in water
Mobile phase B: 0.055% TFA in acetonitrile
Temp.:

35 °C

Wavelength:

214 nm

Injection vol.:

36 µL

Run time:

65 minutes

Gradient:

Time Flow %A %B Curve

0.96 100 0

9.9 0.96 85 15 6

42.9 0.96 64 36 6

45.0 0.96 40 60 11

50.0 0.96 100 0 11

83.0 0

100 0

11

6

8

10

12

14

16

18

20

22

24

26

28

30

32

34

36 min

0.00

0.10

0.20

0.30

0.40

0.50

AU

T1

T13-

T14

T14

T4

T9

-T

10

T10

T19C

T8

T15

T5

T19

T12-

T13

T12

Column:

XBridge BEH130 C

18

, 4.6 x 150 mm, 3.5 μm

Part number:

186003569

Mobile phase A: 0.055% TFA in water
Mobile phase B: 0.055% TFA in acetonitrile
Temp.:

35 °C

Wavelength:

214 nm

Injection vol.:

36 µL

Run time:

65 minutes

Gradient:

Time Flow %A %B Curve

0.96 100 0

9

0.96 85 15 6

39 0.96 64 36 6

43 0.96 40 60 11

58 0.96 100 0 11

67 0

100 0 11

Column

ACQUITY UPLC BEH300 C

18

, 2.1 x 100 mm, 1.7 μm

Part number:

186003686

Mobile phase A: 0.055% TFA in water
Mobile phase B: 0.055% TFA in acetonitrile
Temp.:

35 °C

Wavelength:

214 nm

Injection vol.:

5 µL

Run time:

61 minutes

Gradient:

Time Flow %A1 %B1 Curve

0.2 100 0

6.6 0.2 85 15 6

28.6 0.2 64 36

6

30.0 0.2 40 60

11

40.0 0.2 100 0

11

62.0 0

100 0

11

3

[ Care and Use ManUal ]

Cytochrome c Digestion Standard